Mobiler Quecksilberanalysator für Erdgas
Perfekt geeignet um die Effizienz unterschiedlicher Hg-Absorber (Mercury Removal sytems, MRUs) in der Erdgasverarbeitung zu überwachen.
Jedes Online-System zur Quecksilberbestimmung muss unterschiedlichen Anforderungen bezüglich der Probengaszusammensetzung, der Installation und den Betriebsbedingungen genügen. Wir besitzen die Erfahrung um Erdgassysteme zu liefern, die an Ihre speziellen Bedürfnisse angepasst sind.
- Features & Benefits
- Main applications
- Technical Specifications
- Technical Data
- Approvals & Certifications
- Downloads
- Built-in wheels for easy transport
- Well proven components: UT-3000 Ultratracer and NG Sampling System 3000
- AutoRange function
- Gas alarm for increased safety
- Manual Calibration Set (CalSet) included
- Automatic caibration unit (AutoCal) as an option
- Option: explosion-proof cabinet certified for hazardous zones 1 and 2 (EEX certified)
- Natural gas platforms
- Exploitation of natural gas
- Natural Gas processing plants
Technical Specifications | |
---|---|
Measuring principle | Amalgamation (MI GoldTrap), cold vapor atomic absorption (CVAAS) at 253.7 nm |
UV-Source | Electrodeless low pressure mercury lamp (EDL) |
Stabilization method | Reference beam method |
Optical cell | Fused silica (Suprasil), l: approx. 230mm, heated approx. 45° |
Measured component: | Total Gaseous Mercury (TGM) |
Measuring ranges | owest range: 0.1 ng/m3 – 1000 ng/m3 (at 10 l sample gas volume); medium range: 1 ng/m3 – 10 000 ng/m3 (at 1 l sample gas volume); screening mode: 10 ng/m3 – 30 000 ng/m3 (at 0.1 l sample gas volume) |
Detection limit | 1 pg absolutely or 0.1 ng/m3 |
Linearity over the 0.2 to 15000 ng/m3 range | 3% or better |
Reproducibility | 5% or better |
Technical data | |
---|---|
Signal outputs | USB/RS232 (detector) |
Status output | 3 x pairs of relay contacts (dry contacts) |
Max. sample temperature | 80 °C |
Operating temperature | 0 °C … 40 °C |
Power supply | 230 VAC / 50 Hz (110 VAC / 60 Hz option) |
Power consumption | System: 425 VA; (UT-3000: 250 VA max); (NG Sampling System 3000: 175 VA) |
Dimensions (W x H x D) | 72 x 78 x 78 cm |
Weight | approx. 50 kg |
- Our products are manufactured according to the ISO 9001 quality regulations.
- CE approved